FDA’s Quality Management System Regulation Folds ISO 13485 into US Device Law

From 2 February 2026, 21 CFR Part 820 incorporates ISO 13485:2016 by reference, and FDA inspections changed structure with it.

A technician calibrating a measuring receiver against reference instruments in a precision measurement laboratory.
Photo: MSgt David Miller (Public domain) · via Wikimedia Commons

The FDA’s Quality Management System Regulation took effect on 2 February 2026, amending the device current good manufacturing practice requirements of 21 CFR Part 820 to incorporate ISO 13485:2016 by reference. The rule replaces the Quality System Regulation that had governed US device manufacturing since 1996, and is now titled the QMSR.

The requirement applies to finished device manufacturers intending to commercially distribute devices in the United States. All such manufacturers were required to be compliant by the effective date.

On the same day the agency began using a revised inspection process, set out in the updated Inspection of Medical Device Manufacturers Compliance Program 7382.850. Inspections are now structured around the ISO 13485 framework rather than the former QSR subsystems.

The practical effect

For manufacturers already certified to ISO 13485 this is largely a harmonisation win: one quality system, recognised by more regulators, with the FDA’s framework aligned to what other authorities already use. For organisations that ran a QSR-shaped system, the terminology, documentation structure and risk-management expectations have all shifted.

Control of monitoring and measuring equipment, ISO 13485 clause 7.6, is one of the areas where the language is more explicit than the QSR equivalent. It calls for equipment to be calibrated or verified at specified intervals against measurement standards traceable to international or national standards, with records retained, and where no such standard exists it requires the basis used for calibration or verification to be recorded.

Original source

U.S. Food and Drug Administration

February 2, 2026 · fda.gov

Read the original

This summary was written by Gastronix from publicly reported information. It is not a reproduction of the original article, which remains the property of its publisher. Always check the source before acting on regulatory or safety information.

Share
All news