Step-by-step guides for engineers and technicians: the bench set-ups from ISO 80601, how to run them with LUMA, the GL-1500 test lung and the GR resistances, and how to record the results.
The ten volume-control and pressure-control bench tests from ISO 80601-2-12, with the lung compliance, resistance and ventilator settings for each, and how to measure them with LUMA.
Four bench tests for an anaesthetic workstation from ISO 80601-2-13: oxygen flush flow, flowmeter accuracy, adjustable pressure limit and vapour delivery, with set-ups, limits and how to measure them with LUMA.
How the VentScope Reports tab captures readings from the LUMA analyzer, judges them against acceptance limits, and produces an A4 PDF report with device details, results and an overall verdict.
What a ventilator performance check needs: a gas flow analyzer, a test lung and calibrated resistances, which values to measure, and how to keep a traceable record.
What compliance and resistance mean, how a test lung stands in for a patient, and why a test lung used to verify ventilators needs its own calibration certificate.
What airway resistance values like 5, 20 and 50 cmH2O/L/s mean, why linear resistors are preferred for ventilator tests, and how to choose the right set.