Anaesthetic workstation bench tests to ISO 80601-2-13: oxygen flush, flowmeters, pressure limit and vaporiser

Four bench tests for an anaesthetic workstation from ISO 80601-2-13: oxygen flush flow, flowmeter accuracy, adjustable pressure limit and vapour delivery, with set-ups, limits and how to measure them with LUMA.

5 min readUpdated 23 October 2026← All guides

ISO 80601-2-13 (anaesthetic workstations) sets requirements that can be checked on the bench with a flow analyzer and a test lung. This guide describes four of them, with the set-up, the limit and what to read on the LUMA. Each is based on ISO 80601-2-13:2022 and names its clause. The clause fixes the limit; the test points below (which settings to try) are our choice unless we say otherwise.

1. Oxygen flush (clause 201.101.8)

What the standard asks: the flush must deliver a steady 25 to 75 l/min at the common gas outlet, with a control that has one Off position, can be operated with one hand and closes by itself. It must also have no means of supplying another single gas to the fresh-gas outlet.

Set-up: every flowmeter OFF (air, O₂ and N₂O: any gas still flowing is added to the flush and reads too high). Rotate the selector to open circuit and connect its outlet to the LUMA with a 22 mm breathing hose. LUMA on flow, l/min, gas air/O₂, range of at least ±250 l/min.

Test: press and hold the O₂ flush button and read the flow while it is held. It must rise to a steady 25 to 75 l/min. Release it and check that it closes by itself.

Oxygen flush set-up: the flush outlet connected to the LUMA with a 22 mm breathing hose.
Oxygen flush set-up: the flush outlet connected to the LUMA with a 22 mm breathing hose. Click to enlarge

Two parts: the flow is measured. The control's behaviour (one Off position, one hand, self-closing) is observed, and you mark each as conforming or not.

2. Flowmeter graduations and accuracy (clause 201.101.6.1)

What the standard asks: each graduation within ±10 % of the indicated value, between 10 % and 100 % of full scale, discharged into the ambient atmosphere. Scales graduated in l/min, flows of 1 l/min and below written as ml/min or as decimal l/min with a leading zero, and the method consistent across the system.

Set-up: the open-circuit outlet connected to the LUMA with a 22 mm hose, and the LUMA outlet left open to atmosphere (the clause measures the flow discharged into the room, so no test lung downstream). LUMA on flow, l/min, with its gas selection matching the gas under test.

Test: shut every flowmeter, open only the one under test, set the float on a graduation, let it settle and read the LUMA. The indicated value is the reference. On a 10 l/min flowmeter, the points we use are 1, 2, 4, 6, 8 and 10 l/min on O₂ and on N₂O. The 0.5 mark sits below 10 % of full scale, which the clause does not cover. Change the analyzer's gas selection between O₂ and N₂O, because a flow analyzer reads differently for each.

Flowmeter accuracy set-up: the flowmeter outlet to the LUMA, discharging to atmosphere.
Flowmeter accuracy set-up: the flowmeter outlet to the LUMA, discharging to atmosphere. Click to enlarge

3. Adjustable pressure limit, the APL valve (clauses 201.102.2.2 b and 201.105.2.2 b)

What the standard asks: the pressure at the patient connection port must not exceed the set limit by more than 15 % or 10 hPa (10 cmH₂O), whichever is greater.

Set-up: the LUMA at the patient connection port with the test lung on its outlet. LUMA on pressure, hPa. Select manual or spontaneous ventilation and fit the reservoir bag.

Test: set the APL to a value, let fresh gas raise the pressure until the valve opens and the reading settles, and read the pressure. Repeat for each setting (we use 10, 20 and 40 cmH₂O). Note that "whichever is greater" means the allowance at a setting of 100 is 15 hPa, not 25. The limit is one-sided: a valve that opens early reads low and still passes, so check by eye that it behaves sensibly.

APL valve set-up: LUMA at the patient connection port with the test lung on its outlet.
APL valve set-up: LUMA at the patient connection port with the test lung on its outlet. Click to enlarge

4. Anaesthetic vapour delivery (clause 201.104.2.2): a partial test

What the standard asks: at every graduation above zero, the delivered concentration within +30 % or −20 % of the setting, or +7.5 % or −5 % of the maximum setting, whichever is greater. At Off, Standby or 0 it must not exceed 0.1 % volume fraction.

What we do here: the standard's own test drives the vaporiser against a back-pressure fluctuation of (20 ± 3) hPa, and repeats the sweep at 8 l/min against (50 ± 4) hPa. The steady-flow run described here does neither. It checks accuracy, not conformity with the clause, and a report should not say otherwise.

Run: hold the workstation, analyzer and agent at (20 ± 3) °C for at least 3 hours. Fill the vaporiser to about half its usable volume and let it stand for 45 minutes. Connect the Masimo AX+ gas module, which measures the agent concentration, to the LUMA at the fresh-gas outlet with nothing downstream that absorbs agent or leaks. Set the fresh gas flow to (2 ± 0.2) l/min with the outlet open. Read the concentration with the dial at Off (it must not exceed 0.1 %), then at each setting in turn (for a 0 to 5 % dial: 0.2, 0.6, 1, 2, 4 and 5 %) and let each reading settle.

Vapour delivery set-up: the gas analyzer at the fresh-gas outlet, vaporiser set to each graduation in turn.
Vapour delivery set-up: the gas analyzer at the fresh-gas outlet, vaporiser set to each graduation in turn. Click to enlarge

Recording the results

All four tests produce a table of readings and a verdict. In VentScope the Reports tab captures the readings from the analyzer, applies these limits and prints them in a PDF report. See "Recording and reporting test results in VentScope".

Instruments: LUMA gas flow analyzer and the GL-1500 test lung.