ISO 80601 bench tests

Recording and reporting ventilator and anaesthesia machine tests in VentScope

How the VentScope Reports tab captures readings from the LUMA analyzer, judges them against acceptance limits, and produces an A4 PDF report with device details, results and an overall verdict.

3 min readUpdated 23 October 2026← All guides

A bench test is only useful if you can show it later. VentScope's Reports tab is built for the tests in ISO 80601-2-12 and ISO 80601-2-13: it takes each reading from the analyzer, compares it with the acceptance limit, and builds a report you can file.

Nothing is typed in

Every reading comes from the LUMA. You do not copy numbers from the ventilator's display, and the ventilator's own settings are taken as correct. For each test the Reports tab shows the bench set-up drawing, the settings to apply and a table of the readings to capture, such as tidal volume, PEEP, oxygen concentration, airway pressure, flow or vapour concentration.

For tests where you turn a control between readings, such as the adjustable pressure limit or the flowmeter, each setting has its own capture button, so one reading is never stamped across several lines. For flow readings the oxygen flush captures the highest flow over the last 3 seconds while you hold the button, and the flowmeter test uses the average flow, because a settled flowmeter gives a steady value.

One of the bench drawings the Reports tab shows beside the readings: LUMA at the patient connection port, resistance and test lung.
One of the bench drawings the Reports tab shows beside the readings: LUMA at the patient connection port, resistance and test lung. Click to enlarge

Each reading is judged

The acceptance limit appears next to the reading, with its unit and where it comes from.

  • Limits set by the standard, such as the oxygen flush flow of 25 to 75 l/min, are fixed.
  • Limits the manufacturer discloses, such as the bias and linearity error of volume and pressure delivery in ISO 80601-2-12, are editable. They start as the standard's examples, and you replace them with the figures from the instructions for use of the ventilator under test.
  • Observed items, such as "the flush closes by itself", are marked conforms or does not conform by you. A test with any item still unassessed stays pending and cannot be accepted, because a statement of conformity cannot be made from a partial assessment.

A reading that passes is marked as within the limit; one that fails is marked, and a failed item fails the whole test.

The PDF report

Export PDF Report builds an A4 document. Its header carries:

  • the date and the operator;
  • the device under test (model and serial number);
  • the flow analyzer (model, serial number and calibration due date);
  • the overall verdict.

Each accepted row then prints its bench drawing, the lung and ventilator conditions, and the measurement table. Only accepted rows are included. The report is light-only by design, because it is meant for paper and for filing.

What it does not do

It does not make a ventilator conform to a standard, and it does not replace reading the standard. Some limits depend on the manufacturer's disclosure, and some tests are simplified: the vapour delivery test runs at steady flow without the back-pressure the clause specifies, so a pass is an accuracy result, not a conformity claim. The Reports tab says so on the test.

See also: the VentScope page, and the guides to the ISO 80601-2-12 inflation tests and the ISO 80601-2-13 workstation tests.