When you test a ventilator on a test lung, the lung gives the compliance. The resistance has to come from somewhere else: a small calibrated part placed in the breathing circuit that makes the gas work a little harder to pass. That part is a flow resistance.
What "linear" means
A linear resistance has the same resistance at every flow within its range: double the flow and the pressure drop doubles. The alternative, a non-linear (for example parabolic) resistance, changes with flow. Reviews of ventilator test models generally prefer linear resistances, because the result does not depend on which flow the ventilator happens to be delivering at that moment.
What the values mean
The value is written in cmH₂O/L/s: the pressure, in cmH₂O, needed to push one litre per second through the part. A resistance of 5 is a low, adult-like value; 20 and 50 stand for obstructed airways, and the highest values model severe obstruction or very small airways.
Resistance is quoted with a flow range, because the part only behaves as labelled inside it. The Gastronix GR Series lists the usable flow for each value:
| Model | Resistance (cmH₂O/L/s) | Flow range (L/min) |
|---|---|---|
| GR-5 | 5 | 0–120 |
| GR-10 | 10 | 0–100 |
| GR-20 | 20 | 0–60 |
| GR-50 | 50 | 0–30 |
| GR-100 | 100 | 0–20 |
| GR-200 | 200 | 0–10 |
The higher the resistance, the lower the flow it is meant for, so choose the value that matches the test you are running.
Choosing a set
- Adult ventilator testing usually needs the low and medium values (5 to 50).
- Neonatal and paediatric ventilators run at much lower flows, where the higher values (50 to 200) apply.
- A complete verification uses several values with several compliances, so most laboratories buy the full set.
Each GR resistance is serialised and delivered with its test data, so you can show which part was used in a check. You can buy a single resistance, choosing the value and the quantity, or the full set of six.
Using them with a test lung
Fit the resistance between the ventilator and the test lung, set the ventilator, and read the result on your gas flow analyzer. The three compliances of the lung and the six resistances give eighteen combinations.
See the Webstore for prices in US dollars, or write to customerservice@gastronixlab.com.