A ventilator cannot be tested into thin air. It needs something to push gas into that behaves in a known way. That is the job of a test lung: a mechanical model of a patient's lungs.
Two numbers describe a lung
Compliance is how easily the lung stretches. It is the volume that goes in for each unit of pressure, written in mL/cmH₂O (or L/kPa). A stiff lung has low compliance: a lot of pressure for a little volume. A soft lung has high compliance.
Resistance is how much the airways resist the flow of gas. It is the pressure needed to push a certain flow through, written in cmH₂O/L/s. Narrow or blocked airways mean high resistance.
A ventilator that is accurate with one combination of compliance and resistance is not automatically accurate with another, which is why verification uses several.

What a good test lung gives you
- Known values. The compliance printed on the lung is what you rely on when you judge the ventilator. If the lung is wrong, the check is wrong.
- Several settings. The GL-1500 test lung has three compliances: 0.5, 0.2 and 0.1 L/kPa (about 50, 20 and 10 mL/cmH₂O), each with a stated ±10 % at a stated tidal volume.
- Standard connections. An ISO 22 mm female connection fits standard breathing circuits. Our datasheet lists ISO 80601-2-12, ISO 80601-2-72 and ISO 10651 for the GL-1500.
- Resistance on demand. Test resistances are added in front of the lung. The GR Series gives six calibrated values from 5 to 200 cmH₂O/L/s.
Why the certificate matters
Makers of reference instruments make the same point: a test lung used to check ventilators should come with a calibration certificate, because many general-purpose lungs are not built for the values used in international testing. Without a certificate you cannot say how far off the lung is, and you cannot show an auditor that your check is traceable.
A certificate with full data lists the measured values, not only a pass. Every Gastronix instrument is delivered calibrated with a calibration certificate with full data, and the lung can go back to Gastronix for recalibration.
Getting started
The Lung Simulator Kit combines the test lung with the resistances. See the Webstore for prices in US dollars, or write to customerservice@gastronixlab.com.